Details
-Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
-Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F)
-MP Controller, 2 programs with 10 sections each or alternatively switch over to 1 program with 20 sections
- -The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections
- -Integrated week programm timer with real time function
- -Adjustable time functions and programmable temperature ramp function via program editor
-Digital temperature setting with an accuracy of one degree
-Adjustable fan speed
-Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch)
-Elapsed time indicator
-Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
-RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
-Units up to 115 liters are stackable
-2 chrome-plated racks included
-BINDER test certificate
FP 400 |
|
Exterior dimensions |
|
Width (mm/inch) |
1234 / 48.6 |
Height (inclusive feet) (mm/inch) |
1022 / 40.2 |
Depth (mm/inch) |
765 / 30.1 |
plus door handle, I-panel and exhaust duct (mm/inch) |
105 / 4.1 |
Wall clearance rear (mm/inch) |
100 / 3.9 |
Wall clearance side (mm/inch) |
160 / 6.3 |
Exhaust duct outer- Ø (mm/inch) |
52 / 2.1 |
Steam space volume (l/cu.ft.) |
498 / 17.6 |
Number of doors |
2 |
Interior dimensions |
|
Width (mm/inch) |
1000 / 39.4 |
Height (mm/inch) |
800 / 31.5 |
Depth (mm/inch) |
500 / 19.7 |
Interior volume (l/cu.ft.) |
400 / 14.3 |
Racks, chrome-plated (number standard/max.) |
2 / 10 |
Load per rack (kg/lbs.) |
35 / 77 |
Permitted total load (kg/lbs.) |
90 / 199 |
Weight of the unit (empty) (kg/lbs.) |
145 / 320 |
Temperature data |
|
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) |
300 / 572 |
Temperature variation 1) |
|
at 70 °C (± °C) |
1 |
at 150 °C (± °C) |
2.5 |
at 300 °C (± °C) |
4.8 |
Temperature fluctuation (± °C) |
0.3 |
Heating up time 2) |
|
to 70 °C (Min.) |
18 |
to 150 °C (Min.) |
35 |
to 250 °C (Min.) |
60 |
Recov. time after door was opened for 30 sec. 2) |
|
at 70 °C (Min.) |
2 |
at 150 °C (Min.) |
17 |
at 300 °C (Min.) |
21 |
Air change data |
|
Air change |
|
at 70 °C (x/h) |
17 |
at 150 °C (x/h) |
18 |
at 300 °C (x/h) |
16 |
Electrical data |
|
Housing protection acc. to EN 60529 |
IP 20 |
Nominal voltage (±10 %) 50 / 60 Hz (V) |
400 / 3 N~ |
Nominal power (kW) |
3.4 |
Energy consumption |
|
at 70 °C (W) |
520 |
at 150 °C (W) |
1200 |
at 300 °C (W) |
2340 |
Additional Information
Brand | Binder |
---|